Coating thickness gauge for electroplating and electroless coatings XUL works on x-ray technology for measuring coating thickness. It is known for its compact design and high performance. Our coating thickness gauge is extensively used for measurements on electronic components and small parts like nuts, bolts and screws. It is equipped with primary filters and electrically changeable apertures. This is widely demanded in different industries such as watch and jewellery industries.

Applications

  • Measurements on small parts like screws, bolts and nuts
  • Measurements on contacts and electronic components
  • Determining of the composition of electroplating baths

General Specification

Intended use:Energy dispersive x-ray fluorescence measuring instrument (EDXRF) for material analysis and coating thickness measurement.

Element range:Chlorine (17) to Uranium (92) – up to 24 elements simultaneously with option WinFTM® BASIC.

Design:Bench top unit with upwards opening hood

Measuring direction:Bottom up

X-Ray Source

X-ray tube :-

XUL 210, XUL 220:Micro-focus tungsten tube with beryllium window

XULM 240:Micro-focus tungsten tube with beryllium window

High voltage:Three steps: 30 kV, 40 kV, 50 kV

Aperture (Collimator):

XUL 210, XUL 220:Ø 0.3 mm (11.8 mils), optional slot Ø 0.3 x 0.05 mm (11.8 x 2 mils)

XULM 240:-4 x changeable

Standard (523-440):Ø 0.1 mm (3.9 mils); Ø 0.2 mm (7.9 mils); 0.05 x 0,05 mm (2 x 2 mils);

Ø 0.1 mm (3.9 mils); Ø 0.2 mm (7.9 mils); 0.05 x 0,05 mm (2 x 2 mils);

Optional (523-366):Ø 0.1 mm (3.9 mils); Ø 0.2 mm (7.9 mils); Ø 0.3 mm (11.8 mils);

0.3 x 0.05 (11.8 x 2 mils)

Optional (524-061):Ø 0.1 mm (3.9 mils); Ø 0.2 mm (7.9 mils); 0.3 x 0.05 mm (11.8 x 2 mils);

0,05 x 0,05 mm (2 x 2 mils)

Measurement spot: Depending on the measuring distance and on the aperture, the actual

measurement. Spot size is shown in the video image.

Smallest measurement spot:

XUL 210 and XUL 220: approx. Ø 0,51 mm (20 mils)

XULM 240: approx. Ø 0,1 mm (3.9 mils)

X-Ray Detection

X-ray detector:Proportional counter tube

Absorber:XULM 240 only: optional cobalt or nickel absorber

Measuring distance:0 … 25 mm (0 … 1 in)

Distance compensation with patented DCM method for simplified measurements at varying distances. For particular applications or for higher demands on accuracy an additional calibration might be necessary.

Sample Alignment

Sample positioning:Manually

Video microscope:High-resolution CCD color camera for optical monitoring of the measurement location along the primary beam axis, Crosshairs with a calibrated scale (ruler) and spot-indicator,Adjustable LED illumination.

Zoom factor:Digital 1x, 2x, 3x, 4x

Evaluation Unit

Computer:Windows®-PC

Software:Standard: Fischer WinFTM® LIGHT

Optional: Fischer WinFTM® BASIC, PDM®, SUPER

Sample support stage

Sample Stage


XUL 210

XUL 220,XULM 240

Design

Fixed sample support

Manual operable X/Y-stage

Usable sample placement area

310 x 320 mm (12.2 x 12.6 in)

310 x 320 mm (12.2 x 12.6 in)

Maximum travel XY

-

30 (X) x 40 (Y) mm
1.2 (X) x 1.6 (Y) in

Max. sample Weight

13 kg (29 lb)

2 kg (4.4 lb)

Max. sample Height

174 mm (6.8 in)

174 mm (6.8 in)


Electrical Data

Power supply:AC 115 V or AC 230 V 50 / 60 Hz

Power consumption:max. 120 W, without evaluation PC

Protection class:IP40

Dimensions

External dimensions:Width x depth x height [mm]: 403 x 588 x 444 mm (15.9 x 23.2 x 17.5 in)

Weight:Approx. 45 kg (100 lb)

Environmental Conditions

Operating temperature:10 °C – 40 °C / 50 °F – 104 °F

Storage/Transport0 °C – 50 °C / 32 °F – 122 °F

temperature:

Admissible air humidity:≤ 95 %, non-condensing

Standards

CE approval:EN 61010

X-Ray standards:DIN ISO 3497 and ASTM B 568